Skip to main content

Raman Renishaw specrometer INVIA was purchased specifically having in mind detection of microplastic particles (1 micron to 5 mm in size) using special Particle Analyzer software installed on it. However, this does not limit its application in materials science, chemistry and biology as well. Vibrational spectra can be obtained in the 100-3500 cm-1 (for 785 nm laser) and 100 – 5000 cm-1 (457, 532 nm lasers).

General instrument information

Short name
Raman INVIA
Inventory number
49409
Category
srednja (od 55.000 do 400.000 EUR)
Applications
analitička kemija , identifikacija i kvantifikacija kemijskih spojeva u uzorcima , mineralogija , analiza uzoraka
Standalone or bound
samostalan
Equipment condition
potpuno funkcionalan
Disciplines
atomska i molekulska fizika te optika , fizika kondenzirane tvari , mineralogija i petrologija , fizikalna kemija , analitička kemija
Year of manufacture
2021
Acquisition date
10.03.2022
Financing body
Europska unija
External link

Characteristics

Model
Renishaw INVIA
Manufacturer
Renishaw
Operating and measuring scope
100-3500 cm-1 (for 785 nm laser) and 100 – 5000 cm-1 (457, 532 nm lasers)
Principle of work
dispersive spectrometer
Detailed technical characteristics
It has Renishaw CCD Centrus detector and Andor EMCCD camera with 1600x200 pixels and three laser excitation wavelengths (457, 532 and 785 nm). Accompanying Leica DM2700 microscope facilitates obtaining images in bright field or dark field environment. In addition there is a fluorescence kit for obtaining images in following spectral intervals: blue (450 – 480 nm), green (510 - 550 nm), blue-violet (400 - 440 nm) and UV (330-390 nm). The grating with 2400 grooves per mm facilitates spectral resolution better than 0.5 cm-1 and is used with 532 and 457 nm lasers, while the grating with 1200 grooves/mm is suitable for 785 nm excitation. By recording spectra at particle positions only, Particle Analyzer software enables user to perform swift spectral database search and determination of the particles' chemical composition. Microscope spatial resolution is better than 0.3 microns in the x-y plane and 2 microns in the z direction.
Additional equipment
Particle Analyzer software, fluorescence kit for imaging
Purchase price
273060.59 EUR
Portability
No
Remote work
No

Terms of use

User instructions
only authorised personnel trained by Renishaw
Usage method
korištenje uz pomoć ovlaštene osobe