HPLC je oblik kromatografije na stupcu koji se često koristi u analitičkoj kemiji. Ponekad se u literaturi može naći i naziv "Tekućinska kromatografija pod visokim tlakom" (engl. High pressure liquid chromatography - HPLC). HPLC se koristi za razdvajanje komponenti iz smjese na osnovi kemijskih interakcija između tvari koja se analizira i stacionarne faze u stupcu.

Opći podaci o instrumentu

Skraćeni naziv instrumenta
HPLC
Inventarni broj
0030862, 0308620, 0308621
Kategorija
sitna (< 26.544,56 EUR / 200.000 kn)
Vrsta instrumenta
kromatograf
Samostalan/vezan
samostalan
Stanje opreme
potpuno funkcionalan
Discipline
Kemija
Vanjski link za kapitalnu opremu

Karakteristike

Model
Varian ProStar
Proizvođač
Varian Inc. (danas Agilent Technologies)
Princip rada i mjerna tehnika
Gradient pumps offer selectable one-, two-, or four-stroke proportioning is available to optimize gradient precision on long, shallow gradients while minimizing gradient delay volume on steep, fast gradients. Reliable results -- Intelligent design, including electronic flow and in-line, high-speed proportioning valves, assures accurate and reproducible results. GLP compliance -- Run-and-error logs maintain a complete audit trail of operating parameters, set points, errors, and faults. Built-in automations -- Up to five methods can be built and stored at the pump. Choose the appropriate method with a single keystroke.
Detaljne tehničke karakteristike
;Technical Data Maximum operating pressure: 6000 psi (410 atmospheres) Flow rate range: 0.01 to 10.0 mL/min in 0.01 mL/min steps between 0.00 and 1.00 mL/min, and 0.1 mL/min steps between 1.0 and 10.0 mL/min Flow precision: 0.3% RSD at 1.0 mL/min MeOH/water Flow accuracy: ± 0.5% at 1.0 mL/min isopropyl alcohol Compositional accuracy: ± 0.5% absolute Compositional precision: ± 0.1% absolute Delay volume: 700 µL + 1 µL/ATM pressure Pressure pulsation: Less than 2% at 150 atmospheres (1 mL/min with 100% isopropyl alchohol) Proportioning modes: 1, 2, or 4-stroke (user selectable) Mixing: High pressure, static or dynamic mixing modes Display: 40 Characters, double-line, back-lit LCD
Prenosivost
Ne
Rad na daljinu
Ne

Lokacija

Adresa
Bijenička 54
Krilo/Kat/Soba
IV.K/1.kat/Soba 11c